OpticalFit Manual 18.06 Table of contents Spectrum plot « Last page Next page »

Smoothing

The ‘Smoothing FWHM’ window

The user may apply Gaussian smoothing to noisy spectra using the ‘Spectrum→Smoothing...’ menu option, as below.

The 'Spectrum→Smoothing...' menu option.
The ‘Spectrum→Smoothing...’ menu option.

The ‘Spectrum→Smoothing...’ menu option opens the ‘Smoothing FWHM’ window (shown below), in which you may enter the full-width at half maximum (FWHM) of the smoothing Gaussian function to be applied to the spectrum.

Using the 'Smoothing FWHM' window.
Using the ‘Smoothing FWHM’ window.

Please note:

Smoothing units

Note that the units for the full width at half maximum of the smoothing Gaussian function is displayed in the title of the ‘Smoothing FWHM’ window (“meV” in the above example). If you would like to specify the FWHM in different units, you may do so by typing the units after the value. OpticalFit supports smoothing units of channels (“ch”), electron volts (“eV”, with scalar prefixes, e.g. “meV”, “keV”), nanometres (“nm”), and inverse centimetres (“cm-1”).

Smoothing multiple spectra

If multiple spectra are loaded (see loading multiple spectra), but you don't wish to smooth them all equally, you may individually set the smoothing FWHM for each spectrum using the ‘Smooth FWHM’ column in the ‘Spectrum options’ window (shown below), which is opened using the ‘Spectrum→Spectrum options...’ menu option.

Smoothing using the 'Spectrum options' window.
Smoothing using the ‘Spectrum options’ window.

The ‘Smooth FWHM’ column in the ‘Spectrum options’ window behaves in the same way as the ‘Smoothing FWHM’ window (see above), except that each spectrum may have a different FWHM value (in different units, if desired), or may have no smoothing applied (i.e. FWHM not set).

© Copyright 2008-2018 CSIRO. All rights reserved.