Sylvite CL/SEM spectrum, 20 keV e- beam, CSIRO 2016
Measurement spectrum
Loading...
Measurement details
| Description | CL/SEM spectrum of Sylvite |
| Reference | Otherwise unpublished luminescence spectra measured by the CSIRO Mineral Resources Microbeam Laboratory, Clayton https://research.csiro.au/microbeam-lab/ |
| Material | Sylvite [ KCl ] |
| Specimen temperature | Ambient |
| Analysis technique | Cathodoluminescence (CL), using an Electron Probe Micro-Analyser (EPMA) |
| Instrument | JEOL 8500F |
| Probe | 20 keV electron beam |
| Beam current | 10 nA |
| Probe diameter | 20 μm |
| Channel count | 1024 |
| Wavelength range | 199.5 … 970.546 nm |
| Intensity range | -0.0828 … 86.2 counts |
| Integral | 22289 counts |
| Data filter | Savitzky-Golay, 3rd order, ±20 channels |
| Lines | Unspecified 2.6 eV (100%), 2 eV (47%) |
Specimen details
| Name | P-0182 KCl |
| Owner | CSIRO |
| Locality | Saxony-Anhalt, Germany |
| Related specimen | M 3080 [Museums Victoria], Saxony-Anhalt, Germany |
| Other measurements | Sylvite CL/SEM spectrum, e- beam, CSIRO 2014 |
Composition
| Element | Atoms | At% | Wt% |
|---|---|---|---|
| K | 1 | 50 | 52.4455 |
| Cl | 1 | 50 | 47.5545 |
Measurement image
CL/SEM spectrum of Sylvite
© (C) 2023 CSIRO Mineral Resources
