Fluocerite EDS spectrum, 15 keV e- beam, CSIRO 2024
Measurement details
Description | EDS spectrum of Fluocerite [(Ce,La)F3] in specimen 'S-24-014' (CSIRO) |
Reference | CSIRO 2024 |
Date measured | 2024-04-09 |
Material | Fluocerite [ (Ce, La)F3 ] |
Specimen | S-24-014 Granite [CSIRO - Microbeam Lab client specimens], Benambra, Shire of East Gippsland, Victoria, Australia |
Technique | Energy dispersive x-ray spectrometry (EDS) |
Instrument | JEOL JXA-8530F |
Probe | 15 keV electron beam |
Beam current | 16 nA |
Probe diameter | 1 μm |
Detector name | 2× Bruker XFLASH |
Takeoff angle | 40° |
Max. throughput | 260 kcounts/s |
Real time | 25.6842 s |
Live time | 24.2445 s |
Dead time | 5.60539 % |
Channel count | 4096 |
Energy range | -477.427 … 19990.5 eV |
Gain | 4.99828 eV/channel |
Intensity range | 15937 counts |
Integral | 1.0479×106 counts |
Related measurement | Fluocerite CL spectrum, 15 keV e- beam, CSIRO 2024 (co-measured) |
Composition
Element | Atoms |
---|---|
La | x |
Ce | 1-x |
F | 3 |
Measurement spectrum
Loading...